Farklı teknolojiler kullanılarak üretilmiş güncel nikel-titanyum esaslı döner alet sistemlerinin şekillendirme etkinliklerinin incelenmesi

dc.authorid0000-0003-1622-2698en_US
dc.authorscopusid57212222958en_US
dc.authorwosidCWR-0350-2022en_US
dc.contributor.authorIşık, Vasfiye
dc.contributor.authorErsev, Handan
dc.date.accessioned2024-06-07T06:00:52Z
dc.date.available2024-06-07T06:00:52Z
dc.date.issued2024en_US
dc.departmentFakülteler, Diş Hekimliği Fakültesi, Klinik Bilimler Bölümü, Endodonti Ana Bilim Dalıen_US
dc.description.abstractGİRİŞ ve AMAÇ: Bu çalışmanın amacı farklı termomekanik teknolojiler kullanılarak üretilmiş 4 adet Nikel-titanyum esaslı döner alet sisteminin; ProTaper NEXT (Dentsply Maillefer, Ballaigues, İsviçre), BT-RaCe (FKG, La Chaux-de-Fonds, İsviçre), K3XF (SybronEndo, West Collins, Kaliforniya, ABD) ve HyFlex CM (Coltène-Whaledent, Altstatten, İsviçre), şekillendirme etkinliklerini karşılaştırmaktır. YÖNTEM ve GEREÇLER: S-şekilli kanal içeren reçine bloklardan (n=20) işlem öncesi ve sonrası dijital fotoğraflar; ileri derecede eğimli kanala sahip çekilmiş dişlerden (n=20) şekillendirme öncesi ve sonrası standart yöntemle dijital radyografiler alınmıştır. Alınan fotoğraf ve radyografiler bilgisayar yazılım programı yardımı ile çakıştırılarak ölçümler yapılmış ve transportasyon miktarı ile merkezde kalma oranları hesaplanmıştır. İstatistiksel değerlendirmede varyans analizi ve Kruskal Wallis testi ve post-hoc Tukey ve Dunn’s çoklu karşılaştırma testleri kullanılmıştır. Sonuçlar anlamlılık p<0,05 düzeyinde değerlendirilmiştir. BULGULAR: Reçine bloklarda, kuronal eğimde K3XF merkezde kalma ve transportasyon miktarı açısından en başarılı sonuçları vermiştir (p<0,05). K3XF sistemini sırasıyla HyFlex CM ve ProTaper NEXT sistemleri izlemiştir ancak bu iki sistem arasında istatistiksel olarak anlamlı bir fark saptanmamıştır (p>0,05). BT-RaCe sistemi her iki eğimde de en fazla transportasyon oluşturan (p<0,05) sistem olmuştur. Çekilmiş dişlerde, en az transportasyonu K3XF ve HyFlex CM sistemleri oluşturmuştur (p<0,05). Çalışma uzunluğunda en fazla transportasyona BT-RaCe sistemi neden olmuştur. ProTaper NEXT sisteminin kanal düzleşmesi meydana getirdiği saptanmıştır. TARTIŞMA ve SONUÇ: İleri derecede eğimli ve birden fazla eğime sahip kök kanalların şekillendirilmesinde K3XF en uygun sistem gibi görünmektedir. HyFlex CM sisteminin de güvenle kullanılabileceği sonucuna varılmıştır. Esnekliği arttırılmış alaşımlardan üretilen aletlerin zorlu vakaların şekillendirilmesinde daha başarılı sonuçlar verdiği gözlemlenmiştir.en_US
dc.description.abstractINTRODUCTION: To compare the shaping efficacy of 4 rotary instrumentation systems (ProTaper NEXT (Dentsply Maillefer, Ballaigues, Switzerland), BT-RaCe (FKG, La Chaux-de-Fonds, Switzerland), K3XF (SybronEndo, West Collins, CA, USA) ve HyFlex CM (Coltène-Whaledent, Altstatten, Switzerland) manufactured with different processing technologies of nickel-titanium wire. METHODS: Superimposed pre- and postoperative images of simulated S-shaped canals (n=20) and digital radiographs of severely curved root canals (n=20) were used to determine amount of transportation and centering ratio. The data were analyzed using analysis of variance and Kruskal Wallis test and post-hoc Tukey and Dunn’s multiple comparison tests and significance was set at P<0.05. RESULTS: In simulated S-shaped canals, K3XF gave the best results with respect to amount of transportation and centering ratio at the coronal curvature (P<0.05) followed by HyFlex CM and ProTaper NEXT with no significant differences between them. At the apical curvature, K3XF and HyFlex CM presented significantly better centering ratio results than ProTaper NEXT (P<0.05). BT-RaCe showed significantly higher transportation values than the other systems in both curvatures (P<0.05) In extracted teeth, K3XF and HyFlex CM resulted in the least mean canal transportation (P<0.05) BT-RaCe produced the highest amount of transportation at the apical end point of instrumentation. ProTaper NEXT exhibited pronounced canal straightening. DISCUSSION AND CONCLUSION: Under the conditions of this study, K3XF appears to be more suitable to shape multi-curved and severely curved root canals. HyFlex CM showed adequate canal preparation and was also safe. Alloy types with greater flexibility seem to result in better shaping outcome in complex canals.en_US
dc.identifier.citationIşık, V., & Ersev, H. (2024). Farklı teknolojiler kullanılarak üretilmiş güncel nikel-titanyum esaslı döner alet sistemlerinin şekillendirme etkinliklerinin incelenmesi. 7tepe Klinik Dergisi, 20(2), 4-11. https://doi.org/10.5505/yeditepe.2024.64507en_US
dc.identifier.doi10.5505/yeditepe.2024.64507en_US
dc.identifier.endpage11en_US
dc.identifier.issn1307-8593
dc.identifier.issn2458-9586
dc.identifier.issue2en_US
dc.identifier.startpage4en_US
dc.identifier.urihttps://doi.org/10.5505/yeditepe.2024.64507
dc.identifier.urihttps://hdl.handle.net/20.500.13055/717
dc.identifier.volume20en_US
dc.indekslendigikaynakTR-Dizinen_US
dc.institutionauthorIşık, Vasfiye
dc.language.isoenen_US
dc.publisherYeditepe Üniversitesi Diş Hekimliği Fakültesien_US
dc.relation.ispartof7tepe Klinik Dergisien_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectNiti Döner Aleten_US
dc.subjectİleri Derecede Eğimli Kök Kanalıen_US
dc.subjectS-Şekilli Yapay Kanalen_US
dc.subjectŞekillendirme Etkinliğien_US
dc.subjectTransportasyonen_US
dc.subjectNiti Rotary Instrumentsen_US
dc.subjectSeverely Curved Root Canalen_US
dc.subjectSimulated S-Shaped Canalen_US
dc.subjectShaping Abilityen_US
dc.subjectTransportationen_US
dc.titleFarklı teknolojiler kullanılarak üretilmiş güncel nikel-titanyum esaslı döner alet sistemlerinin şekillendirme etkinliklerinin incelenmesien_US
dc.title.alternativeShaping ability of novel nickel-titanium rotary ınstruments manufactured with different thermal treatmentsen_US
dc.typeArticleen_US
dspace.entity.typePublication

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
Farklı teknolojiler kullanılarak üretilmiş güncel nikel-titanyum esaslı döner alet sistemlerinin şekillendirme etkinliklerinin incelenmesi.pdf
Boyut:
673.73 KB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text
Lisans paketi
Listeleniyor 1 - 1 / 1
Kapalı Erişim
İsim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: