CBCT artifact evaluation in a single device: Insights and limitations

Yükleniyor...
Küçük Resim

Tarih

2024

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Marmara University

Erişim Hakkı

info:eu-repo/semantics/openAccess

Araştırma projeleri

Organizasyon Birimleri

Dergi sayısı

Özet

Objective: To classify the types of artifacts in cone-beam computed tomography (CBCT) and to evaluate them according to age and gender. Methods: CBCT images of 1500 patients (766 males and 734 females) aged 5-92 (mean age: 40.89 ± 18.82 years) were retrospectively evaluated and the patients were categorized into 4 age groups: under 20 years old, 20-39 ages, 40-59 and over 60 years old. The types of artifacts encountered in CBCT images were classified. The relationship between the artifact types with age and gender were investigated. Chi-square test was applied to analyze the relationships between variables and distribution of parameters. Results: Of the cases, 284 (18.9%) were under the age of 20, 389 (25.9%) were between the ages of 20-39, 554 (36.9%) were between the ages of 40-59 and 273 (18.2%) were over the age of 60. Moire artifact was observed at the highest rate (100%), while motion artifact was determined at the lowest rate (19.5%), and no ring artifact was detected in the analyzed images. Metallic artifact, metallic artifact removal, streak artifact and presence of dark bands were found to be statistically significant in females (p = .002, p = .001, p = .002 and p = .002, respectively). There was no statistically significant correlation between cupping artifact, metallic artifact, metallic artifact removal, streak artifact, dark band and noise, and stitched artifact (p > .05). Conclusion: Both device and patient-based artifacts in CBCT images should be known, as well as the ways to prevent them.

Açıklama

Anahtar Kelimeler

Artifacts, Metallic Artifact, Streak Artifact, Motion Artifact, Cone-Beam Computed Tomography

Kaynak

Clinical and Experimental Health Sciences

WoS Q Değeri

Q4

Scopus Q Değeri

Cilt

14

Sayı

2

Künye

Yalçın, E. D., & Aslan Öztürk, E. M. (2024). CBCT artifact evaluation in a single device: Insights and limitations. Clinical and Experimental Health Sciences, 14(2), pp. 349-356. https://doi.org/10.33808/clinexphealthsci.1291106